After-sales service |
| ¡¡ |
|
If
any problem occurred during operation,
please contact with local dealer or our
company. We will try the best to offer good
service. We also offer technical support to
our distribution partners. |
|
¡¡ |
|
Trouble shooting for some parts |
|
1. |
Q: |
The chair can not be moved, but the beep that hints for
function is regular and other keys are in good work |
|
¡¡ |
A: |
1/ Bad connection for
relay, or malfunction for wire coil. Clean the
contacting point of relay, or change the relay.
2/ The motor is in malfunction, or the wire is with open
circuit. Check and re-connect the broken circuit, or
change the motor. |
|
¡¡ |
¡¡ |
|
¡¡ |
¡¡ |
|
¡¡ |
¡¡ |
¡¡ |
|
2. |
Q: |
The X-ray viewer does not shine |
|
¡¡ |
A: |
1/ The bulb is in
malfunction. Change the bulb.
2/ Bad connection for wires. Re-connect the wires. |
|
¡¡ |
¡¡ |
|
¡¡ |
¡¡ |
¡¡ |
|
3. |
Q: |
The instrument tray dangles down |
|
¡¡ |
A: |
1/ The weight of article
placed on the tray£¾2Kg. Reduce the weight exerted on the
tray.
2/ The balance spring worn out. Change the balance
spring.
3/ The balance damping worn out. Change the balance
damping. |
|
¡¡ |
¡¡ |
|
¡¡ |
¡¡ |
|
¡¡ |
¡¡ |
|
¡¡ |
¡¡ |
¡¡ |
|
More
information concerning trouble shooting,
please refer to following FAQ technical
documents |
|
¡¡ |
¡¡ |
¡¡ |
|
Documents list |